Remote Control Commands
R&S
®
NGL200/NGM200
118
User Manual 1178.8736.02 ─ 09
Example: Configuring the resistance limit
// ************************************************
// Select the channel
// ************************************************
INST OUT1
// ************************************************
// Set the resistance value
// ************************************************
// selects a channel and sets the resistance
RES 10
// queries the resistance of the selected channel
RES?
// response: 10.000
// ************************************************
// Query the range of the resistance values
// ************************************************
// queries the upper and lower limit of the resistance
RES? MIN
// response: 0.000
RES? MAX
// response: 10000
// ************************************************
// Aactivate the constant resistance mode
// ************************************************
// selects a channel and activate the constant resistance mode
RES:STAT 1
// queries the constant resistance mode
RES:STAT?
// response: 1
[SOURce:]RESistance[:LEVel][:IMMediate][:AMPLitude]
.....................................................118
.......................................................................................... 119
[SOURce:]RESistance[:LEVel][:IMMediate][:AMPLitude]
<resistance>
[SOURce:]RESistance[:LEVel][:IMMediate][:AMPLitude]?
[<Channel list>]
Sets or queries the constant resistance target value.
Parameters:
<resistance>
<numeric value> | MIN | MINimum | MAX | MAXimum | UP |
DOWN | <list>
<numeric value>
Numeric value in the range of 0.000 ohm to 10000 ohms.
MIN | MINimum
Minimum resistance at 0.000 ohm.
MAX | MAXimum
Maximum resistance at 10000 ohms.
UP
Increases resistance by a defined step size.
Configuration Commands