Remote Control Commands
R&S
®
NGL200/NGM200
119
User Manual 1178.8736.02 ─ 09
DOWN
Decreases resistance by a defined step size.
Increment: 0.1 ohms
Default unit: ohms
Parameters for setting and query:
<Channel list>
<list>
Example:
RES? (@1)
Queries the constant resistance target value at channel 1.
Example:
See
Example "Configuring the resistance limit"
[SOURce:]RESistance:STATe
<state>
[SOURce:]RESistance:STATe?
[<Channel list>]
Sets or queries the constant resistance mode.
Parameters:
<state>
0
Deactivates constant resistance mode.
1
Activates constant resistance mode.
Parameters for setting and query:
<Channel list>
<list>
Example:
RES:STAT? (@1)
Queries the constant resistance mode at channel 1.
Example:
See
Example "Configuring the resistance limit"
7.5.6
Combined Setting of Voltage and Current Setting
The
APPLy
subsystem provides a command that enables you to set the current and
voltage of a channel in one step.
The combined voltage and current setting command takes approximately 100 ms, i.e.
longer than the setting of a single value.
APPLy
<voltage> [,<current>][,<output>]
Sets or queries the voltage and current value of the selected channel.
Parameters:
<voltage>
<numeric value> | MIN | MINimum | MAX | MAXimum | DEF |
DEFault
<numeric value>
Numeric value for voltage in the range of 0.000 to 20.050.
MIN | MINimum
Min voltage at 0.000 V.
Configuration Commands