Remote Control Commands
R&S
®
NGL200/NGM200
120
User Manual 1178.8736.02 ─ 09
MAX | MAXimum
Max value for voltage at 20.050V.
DEF | DEFault
Default voltage.
*RST:
1.000
Default unit: V
<current>
<numeric value> | MIN | MINimum | MAX | MAXimum | DEFault
<numeric value>
Numeric value for current in the range of 0.000 to 6.010.
MIN | MINimum
Min current at 0.000 A.
MAX | MAXimum
Max value for current at 6.010 A.
DEF | DEFault
Numeric value for current.
*RST:
1.000
Default unit: A
<output>
OUT1 | OUTP1 | OUTPut1 | OUT2 | OUTP2 | OUTPut2
OUT1 | OUTP1 | OUTPut1
Selects output for channel 1.
OUT2 | OUTP2 | OUTPut2
Selects output for channel 2.
Example:
INST OUT1
APPL 6,2
Sets 6 V and 2 A to output of channel 1.
APPL?
-> 6.000, 2.000
Queries the voltage and current of the selected channel.
READ?
[<Channel list>]
Queries the currently voltage-current measurement pair and returns them separated by
comma.
Parameters:
<Channel list>
<list>
Example:
READ?
-> 1.00,1.004224E-02
Example:
READ? (@1)
Queries the voltage-current measurement pair at channel 1.
Usage:
Query only
7.5.7
Output Setting
The
OUTPut
subsystem contains the commands for activating the output channels.
Configuration Commands