Remote Control Commands
R&S
®
NGL200/NGM200
128
User Manual 1178.8736.02 ─ 09
Example:
SENS:CURR:RANG:AUTO? (@1)
Queries auto range state for current mesurement accuracy at
channel 1.
SENSe:CURRent:RANGe[:UPPer]
<arg0>[, <Channel list>]
SENSe:CURRent:RANGe[:UPPer]?
[<Channel list>]
Sets or queries the current range for measurement. There is a selection of 10 A, 1 A,
100 mA and 10 mA range.
Parameters:
<arg0>
Defines the current range for measurement (10 A, 1 A, 100 mA
and 10 mA).
Default unit: A
Parameters for setting and query:
<Channel list>
<list>
Example:
SENS:CURR:RANG 10
Sets the instrument to the 10 A measurement accuracy.
Refers to datasheet for the measurement accuracy in the 10 A
range.
Example:
SENS:CURR:RANG? (@1)
Queries current range for measurement at channel 1.
SENSe:VOLTage:RANGe:AUTO
<arg0>[, <Channel list>]
SENSe:VOLTage:RANGe:AUTO?
[<Channel list>]
Sets or queries auto range for voltage measurement accuracy.
Parameters:
<arg0>
1
Enables auto range for voltage.
0
Disables auto range for voltage.
Parameters for setting and query:
<Channel list>
<list>
Example:
SENS:VOLT:RANG:AUTO 1
Enables auto range for voltage.
Example:
SENS:VOLT:RANG:AUTO? (@1)
Queries auto range state for voltage measurement accuracy at
channel 1.
SENSe:VOLTage:RANGe[:UPPer]
<arg0>[, <Channel list>]
SENSe:VOLTage:RANGe[:UPPer]?
[<Channel list>]
Sets or queries the voltage range for measurement. There is a selection of 20 V and 5
V range.
Configuration Commands