Remote Control Commands
R&S
®
NGL200/NGM200
142
User Manual 1178.8736.02 ─ 09
Parameters for setting and query:
<Channel list>
<list>
Example:
POW:PROT:LEV? (@1)
Queries OPP value at channel 1.
[SOURce:]POWer:PROTection:TRIPped?
[<Channel list>]
Queries the OPP state of the selected channel.
Parameters:
<Channel list>
<list>
Example:
POW:PROT:TRIP?
Response 1, the OPP is tripped.
Response 0, the OPP is not tripped.
Example:
POW:PROT:TRIP? (@1)
Queries OPP state at channel 1.
Example:
See
Example "Configuring the overvoltage protection"
Usage:
Query only
7.5.12
USB Class Setting
The
Interface
subsystem contains the commands for changes made on the USB
class.
INTerfaces:USB:CLASs
<arg0>
Sets or queries the USB class.
Parameters:
<arg0>
CDC | TMC
CDC
USB CDC connection.
TMC
USB TMC connection.
7.6
Measurement Commands
The
MEASure
subsystem provides commands to query the voltage and current values
of a channel.
......................................................................................... 143
.................................................................................143
.................................................................................144
...................................................................................144
Measurement Commands