Remote Control Commands
R&S
®
NGL200/NGM200
158
User Manual 1178.8736.02 ─ 09
DIO:FAULt:SIGNal
<arg0>
Select digital output fault signal type.
Parameters:
<arg0>
CONStant | PULSe
CONStant
An constant level trigger signal is sent out
PULSe
An output pulse of 100 ms trigger signal is sent out
DIO:OUTPut:SOURce
<arg0>, <arg1>
DIO:OUTPut:SOURce?
<arg0>
Sets or queries the digital output source.
.
Parameters:
<arg1>
OUTPut | TRIGger | FORCed
OUTPut
Selected channel output is used as the digital output source.
TRIGger
Selected channel external trigger signal is used as the digital
output source.
FORCed
Selected output is forced to high level and can be switched by
on page 158 command.
Parameters for setting and query:
<arg0>
1 | 2
Channel selection for the digital output source.
Example:
DIO:OUTP:SOUR 1, OUTP
Channel 1 output is used as the digital output source.
Example:
DIO:OUTPut 1,0
DIO:OUTPut:SOURce 1,FORC
DIO:OUTPut 1,1
Digital I/O pin OUT1 is set to a "high" level
DIO:OUTPut[:STATe]
<arg0>, <arg1>
DIO:OUTPut[:STATe]?
<arg0>
Sets or queries the digital output channel selection.
Parameters:
<arg1>
1 | 0
Enables or disables the digital output state.
Advanced Operating Commands