Additional Basics on Remote Control
R&S
®
NGL200/NGM200
187
User Manual 1178.8736.02 ─ 09
Depending on the value of the read register, you can draw conclusions on the current
status of the device. For example, when the unit operates in constant voltage, the
result of the returned ISUM register is a decimal "2" which corresponds the binary
value of "0000000000000010".
Any part of a status register system can be read by query commands. A decimal value
is returned and represents the bit pattern of the requested register. Each SCPI register
is 16 bits wide and has various functions. The individual bits are independent, i.e. each
hardware status is assigned to a bit number.
Bits 9 to 12 are still "free" resp. unused (always return a "0"). Certain areas of the reg-
isters are not used. The SCPI standard defines only the "basic functions". Some devi-
ces offer an advanced functionality.
Each channel of the power supply is considered as separate "instrument" (SCPI stan-
dard definition). Therefore, e.g. the register
STATus:QUEStionable:INSTrument:ISUMary
of the NGL202 or NGM202 is also
present two times (Isummary1 - Isummary2).
Description of the status register parts
The SCPI standard provides two different status registers:
Figure A-1: Structure of the status:QUEStionable register
Status Reporting System