Remote Control Commands
R&S
®
SMC100A
208
Operating Manual 1411.4060.32 ─ 04
Parameters:
<Action>
0 | 1
0
The contents of the status registers are preserved.
1
Resets the status registers.
*RCL
<Number>
ReCaLl calls up the instrument settings from an intermediate memory identified by the
specified number. The instrument settings can be stored to this memory using the com-
mand
It also activates the instrument settings which are stored in a file and loaded using the
MMEMory:LOAD <number>, <file_name.extension>
command.
*RST
ReSeT
Sets the instrument to a defined default status. The default settings are indicated in the
description of commands.
The command is equivalent to
SYSTem:PRESet
.
Usage:
Setting only
*SAV
<Number>
SAVe stores the current instrument settings under the specified number in an intermedi-
ate memory. The settings can be recalled using the command
number.
To transfer the stored instrument settings in a file, use the command
.
*SRE
<Contents>
Service Request Enable
Sets the service request enable register to the indicated value. This command determines
under which conditions a service request is triggered.
Parameters:
<Contents>
Contents of the service request enable register in decimal form.
Bit 6 (MSS mask bit) is always 0.
Range:
0 to 255
Common Commands