Configuring the R&S TSME6
R&S
®
TSME6
59
User Manual 4900.0033.02 ─ 06
6.2
Interacting with R&S ROMES, R&S NESTOR, and R&S
ViCom
Measurements are performed on the R&S TSME6 via the R&S
ROMES, R&S
NES-
TOR, or R&S
ViCom interface. Depending on the installed technology options, the net-
work is either scanned for individual standards, or globally over all frequency bands.
The following sections briefly describe the basic requirements for interaction between
the R&S TSME6 and the R&S
ROMES, R&S
NESTOR, or R&S
ViCom interface. For
details on using these interfaces see the R&S
ROMES, R&S
NESTOR, and
R&S
ViCom manuals.
6.2.1
Interacting with R&S ROMES
Prerequisites
To perform measurements with the R&S
ROMES application, the following prerequi-
sites apply:
For details on R&S
ROMES, see the manual provided on the software DVD.
●
R&S
ROMES installation (refer to R&S
ROMES manual)
●
R&S
ROMES option on dongle inserted (refer to R&S
ROMES manual)
●
Required technology options installed on the R&S TSME6 (see
"Installing and managing software license keys - "Options""
●
Software options installed on the host PC (see information on the ROMES product
page on the Rohde
&
https://www.rohde-schwarz.com/product/
After the R&S TSME6 has finished booting, start the R&S
ROMES software applica-
tion.
Automatic firmware updates
Each time R&S
ROMES is started, the software automatically compares the firmware
version of the connected R&S TSME6 with the version currently available on the host
PC. If a newer version is available, it is copied to and installed on the R&S TSME6.
Up to three previously installed firmware versions are stored as backups and main-
tained on the R&S TSME6 (see also
Chapter 6.1.1, "Obtaining device information -
on page 47). Thus, it is not necessary to perform firmware updates on
the R&S TSME6 manually.
Interacting with R&S ROMES, R&S NESTOR, and R&S ViCom