Getting Started
R&S
®
TSME6
38
User Manual 4900.0033.02 ─ 06
Table 3-5: LAN LED 2 states and their meaning
LED state
Description
off
No connection
yellow
Physical connection established
yellow, blinking
Identifying connected device
DC IN connector
The DC IN connector is required for the DC power supply (10-28
V, max. 1.8
A). For
Chapter 3.1.4, "Connecting to power"
RF IN connector (50
Ω
)
The optional multi-band RF antenna (700
MHz to 2.6
GHz) or the device providing the
RF signal is connected to the instrument's RF INPUT via a cable equipped with an
appropriate connector (SMA
female, 50
Ω
input impedance, VSWR
type
2.0).
Do not overload the maximum allowed input of 20 dBm. Non-compliance will destroy
the input mixer.
Depending on the intended use, the respective valid regulations regarding lightning
protection of the antennas and regarding vehicle installation must be observed during
installation.
Pwr./State LEDs
Table 3-6: POWER and STATE LED states and their meaning
STATE LED
POWER LED
Meaning
off
off
no power supply connected at DC IN
power supply off
power supply < 10
V
off
yellow
standby
off
green, blinking (2
Hz)
FGPA configuration in progress
red
(up to 5 seconds
during startup)
green
FPGA configuration finished, preparing for start
off
green
R&S
TSME6 ready, not connected
green
green
connected
green, blinking rap-
idly
green
measuring
green, blinking 2
Hz green
Instrument is identified by the software
red, blinking 2
Hz
green
temperature warning (controller board temperature =
75°
C ... 80°
C)
red (continuous)
green
temperature error (controller board temperature above
80°
C)
Instrument tour