41
strength to withstand the load they will be exposed to. If the bolts are too long, the
wheelchair’s
batteries or other wheelchair components may be damaged. If the bolts have been made too
short, replace them only with original Dahl bolts #502800.
1.
The bolts are secured with Loctite 222 (or an equivalent from another
manufacturer) to secure threads on all bolts, and tightened to torque 16-18
Nm.
2.
Install battery, wiring and fuse. Close the wheelchair again.
⚠
Do not use any bolts other than the original bolts, which are supplied by Dahl Engineering
(Art # 502800, which are of quality 14.9, torx key size 27). Standard M8 bolts with
countersunk heads are not strong enough in the event of a collision.
Carry out a final check by locking the wheelchair firmly in position in the docking station.
Check that the locking plate is correctly locked and verify that all release methods work as
intended. A warning tone will sound if the locking plate has not been inserted all the way into
the docking station.
Description of function for Dahl Docking System
(1)
Dahl Docking Station
(2)
Locking plate and 8 mm spacer plate
(3)
Locking pin
(4)
Red LED
(5)
Green LED
(6)
Control panel
(7) Push button/Release button
(8)
Manual emergency release
(9)
Manual release catch