Controller functions
RES-403
Page 29
Part 3 of 3: Error messages as of October 2005 (warnings)
NOTE:
The specified error messages are initially output as warnings (actual value output jumps back and forth
between two values, alarm LED blinks, alarm relay de-energized). When the "START" signal is activated, the
warning changes to a fault (actual value output no longer jumps back and forth, see bold italic values, alarm
LED lit continuously, alarm relay energized).
Error
code
Act. val.
output
voltage
[V]
Cause
Action if machine
started for first time
Action if machine
already operated, HS
band not changed
111
¯
6.66
«
ª
10
®
Current signal incorrect,
no calibration possible
Fault area
j
,
check configuration
Fault areas
fgh
("loose contact")
112
¯
7.33
«
ª
10
®
Voltage signal incorrect,
no calibration possible
Fault area
i
,
check configuration
Fault areas
fgh
("loose contact")
113
¯
8.00
«
ª
10
®
Current / voltage signals incor-
rect,
no calibration possible
Fault area
ij
,
check configuration
Fault areas
fgh
("loose contact")
114
¯
8.66
«
ª
10
®
Temperature fluctuates,
no calibration possible
Perform
AUTOCAL
and / or
fault areas
fgh
("loose contact")
Perform
AUTOCAL
and / or
fault areas
fgh
("loose contact")
115
Ext. calibration temp. too high,
no calibration possible
Perform
AUTOCAL
with
ext. calibration tempera-
ture
≤
40°C
Perform
AUTOCAL
with
ext. calibration tempera-
ture
≤
40°C
116
Ext. calibration temp. fluctu-
ates,
no calibration possible
Perform
AUTOCAL
with
stable ext. calibration
temperature
Perform
AUTOCAL
with
stable ext. calibration
temperature