Device functions
Page 52
RES-5008
Version 2
Part 3 of 3: Error messages (warnings)
NOTE:
The specified error messages are first output as warnings:
• Actual value output changes between two values
• Alarm-LED blinks slowly
• Alarm relay is not active
After activation of the START signal, the output occurs as a fault
• Actual value output does not change any more, see bold italic values
• Alarm LED lights up continuously
• Alarm relay is active
Error no.
Actual
value
output
Voltage
[V]
Cause
Measure during initial
startup
Measure during
machine operation,
heating element
not changed
10
111
6.66
10
Current signal incorrect,
Calibration not possible
Error range
Check configuration
Error ranges
(“Loose contact”)
11
112
7.33
10
Voltage signal incorrect,
Calibration not possible
Error range
,
Check configuration
Error ranges
(“Loose contact”)
12
113
8.00
10
Voltage/current signal
incorrect,
Calibration not possible
Error ranges
Check configuration
Error ranges
(“Loose contact”)
13
114
8.66
10
Temperature fluctuates,
Calibration not possible
Perform
AUTOCAL
and/or
Error ranges
(“Loose contact”)
Perform
AUTOCAL
and/or
Error ranges
(“Loose contact”)
115
Ext. calibration temperature
too large,
Calibration not possible
Perform
AUTOCAL
with
ext. calibration
temperature ≤40 °C
Perform
AUTOCAL
with
ext. calibration
temperature ≤40 °C
116
Ext. calibration temperature
fluctuates,
Calibration not possible
Perform
AUTOCAL
with
stable ext. calibration
temperature
Perform
AUTOCAL
with
stable ext. calibration
temperature