Electrical components
Wiring circuit
The nursing table comes with control box mounted on the frame. The diagram
shows how to connect the cables with the control box.
All plug-in connections are waterproof, and mounting/dismounting may therefore
be a little difficult. Make sure that all plugs have the correct position before
inserting them into the control unit. The plugs must be pressed home until the O-
ring is no longer visible.
Motor and control box has an IPX6 rating. This means that they are dust-and
waterproof.
The electrical system is designed for intermittent use, which means a maximum duty
cycle of 10%; 2min on / 18min from.
1.
Controlbox
2.
Mains cable 230V
3.
Actuator, connects to 1
4.
Handset
5.
Safety stop
1