Reference Manual
00809-0100-4377, Rev BA
September 2010
Rosemount 752
4-6
LCD TRANSDUCER
BLOCK
This section describes error conditions found in the LCD Transducer Block.
Read Table 4-5 and to determine the appropriate corrective action.
Self Test Procedure for the LCD
The SELF_TEST parameter in the Resource block will test LCD segments.
When running, the segments of the display should light up for about five
seconds.
If your host system supports methods refer to your host documentation on
how to run the
Self Test
method. If your host system does not support
methods than you can run this test manually be following the steps below.
1. Put Resource block into
OOS
(Out of Service).
2. Go to the parameter called
SELF_TEST
and write the value Self test
(0x2).
3. Observe the LCD screen when you are doing this. All of the segments
should light up.
4. Put the Resource block back into
AUTO
.
Table 4-5. LCD Transducer
Block BLOCK_ERR messages
Condition Name and Description
Other
Out of Service
: The actual mode is out of service.
Symptom
Possible Causes
Recommended Action
The LCD displays “DSPLY#INVLID.” Read the
BLOCK_ERR and if it says “BLOCK
CONFIGURATION” perform the Recommended Action
One or more of the display parameters are
not configured properly.
See “LCD Transducer Block" on
page 2-16.
“752” is being displayed or not all of the values are
being displayed.
The LCD block parameter
“DISPLAY_PARAMETER_SELECT is not
properly configured.
See “LCD Transducer Block" on
page 2-16.
The display reads OOS
The resource and or the LCD Transducer
block are OOS.
Verify that both blocks are in
“AUTO,”
The display is hard to read.
Some of the LCD segments may have
gone bad.
See Self Test procedure above. If
some of the segment is bad, replace
the LCD.
Device is out of the temperature limit for the
LCD. (-20 to 85 °C)
Check ambient temperature of the
device.
Summary of Contents for 752
Page 2: ......
Page 4: ......
Page 10: ...Reference Manual 00809 0100 4377 Rev BA September 2010 Rosemount 752 1 4 ...
Page 26: ...Reference Manual 00809 0100 4377 Rev BA September 2010 Rosemount 752 2 16 ...
Page 38: ...Reference Manual 00809 0100 4377 Rev BA September 2010 Rosemount 752 A 4 ...
Page 43: ...Reference Manual 00809 0100 4377 Rev BA September 2010 B 5 Rosemount 752 APPROVAL DRAWINGS ...
Page 44: ...Reference Manual 00809 0100 4377 Rev BA September 2010 Rosemount 752 B 6 ...
Page 45: ...Reference Manual 00809 0100 4377 Rev BA September 2010 B 7 Rosemount 752 ...
Page 46: ...Reference Manual 00809 0100 4377 Rev BA September 2010 Rosemount 752 B 8 ...
Page 47: ...Reference Manual 00809 0100 4377 Rev BA September 2010 B 9 Rosemount 752 ...
Page 48: ...Reference Manual 00809 0100 4377 Rev BA September 2010 Rosemount 752 B 10 ...
Page 59: ......