6
Icons:
Hypertension Risk Indicator: Compares blood pressure against guidelines.
Memory Zone 1: Appears when the monitor is operated in memory zone 1
Memory Zone 2: Appears when the monitor is operated in memory zone 2
Weak Battery Mark: Appears when batteries should be replaced
Pulse Mark: Shows the pulse rate per minute
Irregular Heartbeat (IHB) Detector: The appearance of this symbol signifies that an irregular
heartbeat was detected
Memory Average: Displays average of last 3 readings
Memory Mark: Shows the sequence number of stored measurements
Calibration Indicator: Appears when the monitors is required re-calibration. Re-calibration
recommended.
Measurement Error: Make sure the L-plug is securely connected to the air socket and
measure again. Wrap the cuff correctly and keep arm steady during measurement. If the error keeps
occurring, return the device to your local distributor or service center.
Air Circuit Abnormality: Make sure the L-Plug is securely connected to the air socket on the
side of the unit and measure again. Another possible cause can be due to the short circuit of
the microphone embedded in the cuff. If the errors still occur, return the device to your local distributor
or service center for help.
Pressure Exceeding 300 mmHg: Switch the unit off and measure again. If the error keeps
occurring, return the device to your local distributor or service center.
Data Error: Remove the batteries, wait for 60 seconds, and reload. If the error keeps occurring,
return the device to your local distributor or service center.
Exceeding Measurement Range: Measure again. If the error keeps occurring, return the
device to your local distributor or service center.