SAFETY INFORMATION

 

 

Take all necessary precautions for using any electrical device. Before connecting the electrical supply, 
check to see if the supply voltage is within the range stated at the rating label, and see to it that the 
device be seated firmly. Place the unit in a safe and dry; it must NOT touch the surrounding. Follow 
the safety precautions for chemicals and dangerous materials. If needed, please contact us at Tel.: 
0721/5606-0. 
 

Environmental Conditions 

Ensure the instrument is installed and operated strictly under the following conditions: 

• 

Indoor use only 

• 

RH 

95 % 

• 

75 KPa-106 kPa 

• 

Altitude not to exceed 2000 meters 

• 

4 °C ~ 40 °C operating temperature 

• 

Pollution degree: 2 

• 

Mains supply voltage fluctuations up to ± 10 % of the normal voltage 

 

Avoiding Electrical Shock 

Follow the guidelines below to ensure safe operation of the unit. 
The Roth power supply series has been designed for use with insulated wires, minimizing any 
potential shock hazard to the user. We recommend against the use of uninsulated wires. 
To avoid electrical shock: 
1. In case of spillage, any parts of the cable or power supply must be dried out for a period of time and 

restored to NORMAL CONDITION before the operation.  

2. NEVER connect or disconnect wire leads from the power jacks when the red indicator light of the 

power switch is on. 

3. WAIT at least 5 seconds after stopping a run before handling output leads or connected apparatus. 
4. ALWAYS make sure that your hands, work area, and instruments are 

clean

 and 

dry 

before making 

any connections or operating the power supply. 

5. ONLY connect the power cord to a properly grounded AC outlet. 
 

Avoiding Damage to the Instrument 

1. Do not attempt to operate the device if damage is suspected. 
2. Protect this unit from physical damage, corrosive agents and extreme temperatures (direct sunlight 

etc). 

3. For proper ventilation and safety concerns, keep at least 10 cm of space behind the instrument, and 

at least 5 cm of space on each side. 

4. Use high level of precaution against damaging the unit. 
5. Do not operate the unit out of environmental conditions addressed above. 
6. Do not operate the power supply in high humidity environments (> 95%), or where condensation 

may occur. 

7. To avoid condensation after operating the power supply in a cold room, wrap the unit in a plastic 

bag and allow at least 2 hours for the unit to equilibrate to room temperature before removing the 
bag and operating the unit. 

8. Prior to a

pplying any cleaning or decontamination methods other than manufacturer’s 

recommendation, users should check with the manufacturer’s instruction to see if the proposed 
method will damage the equipment. 

 

Maintenance 

The ROTH power supplies require little maintenance to ensure reliable operation. The housing may 
be cleaned with a dry cloth. 
 

Summary of Contents for MINI

Page 1: ...ully before operating this equipment 1 Read and follow the manual instructions carefully 2 Do not alter the equipment Failure to follow these directions could result in personal and or laboratory hazards as well as invalidate the equipment warranty 3 Use a properly grounded electrical outlet of correct voltage and current handling capacity 4 Disconnect from the power source before maintenance and ...

Page 2: ...ight of the power switch is on 3 WAIT at least 5 seconds after stopping a run before handling output leads or connected apparatus 4 ALWAYS make sure that your hands work area and instruments are clean and dry before making any connections or operating the power supply 5 ONLY connect the power cord to a properly grounded AC outlet Avoiding Damage to the Instrument 1 Do not attempt to operate the de...

Page 3: ...t into a municipal trash bin when this unit has reached the end of its lifetime To ensure utmost protection of the global environment and minimize pollution please recycle this unit Caution Warning Indicates a hazardous situation which if not avoided could result in death or serious injury Max voltage 500 V Max current 400 mA Max watt 120 W Encountering Problems Check the troubleshooting section C...

Page 4: ...multaneously saving both time and valuable bench space With a universal voltage rating the Roth MINI is also designed and constructed to the most rigorous safety standards This makes it an excellent choice for education or personal use FEATURES Compact size Power capacity Constant Voltage or Constant Current operation 1 V step voltage selection 1 mA step current selection Timer No load detection S...

Page 5: ...t Constant V or mA with automatic crossover During constant mode not adjusted parameters are automatically set to maximum Programs Manuell adjustable mode Constant Mode 2 step program editable 30 programs each with up to 6 steps editable Program Mode 2 step program pre assigned Pre assigned programs for different applications Typical running condition Mode Outlets parallel socket pairs 2 4 mm 5 4 ...

Page 6: ... 1 Before operation press MODE to set the parameters of Voltage Current and Time of Step 1 and 2 LED light indicates your current parameter For example if you are setting a value for Voltage the Voltage LED will be lit 2 Two step buttons are available for you to select the step and to check the parameters of voltage current and time of each step User will press STEP 1 and STEP 2 buttons mainly to ...

Page 7: ...example to operate under constant voltage adjust current to max before running constant voltage and vice versa 1 Adjust the output value by pressing the or button 2 After adjusting your constant mode output value switch to the other parameter and set it to maximum 3 Timer Setting After setting the current or voltage output value press MODE to switch to time setting mode then press key or key to ad...

Page 8: ...making sure the tape is removed from the pre cast gel that the buffers are prepared correctly and the recommended volume of buffer is added to the electrophoresis unit Error Message Cause Solution Current exceeds the maximum output for the power supply 400 mA Check if the buffer concentration or molarity is appropriate Excessive buffer concentration or molarity may increase conductivity To clear t...

Reviews: