SI manual – Section: Basic Settings for Commissioning
25
7.4 Limits
NOTE:
Settings and operation must
be verified by electric operation and a
functional test of the actuated valve.
Connect to the actuator as described
in section 7.1. From the Position display
home screen press the
key. The
main menu will be displayed.
Navigate to Settings using the
keys and press
to select.
Y
Y
R
R
Fig. 7.4.1.
The settings menu will be displayed:
Settings
Limits
Stroke Tests
Indication
Control
Security
Defaults
Navigate to Limits using the keys
and press
to select.
The setting first selected to be
changed will require a password to
be entered, refer to section 7.2.
The limit settings are shown below with their factory default values:
Stopped
Close Settings
1 / 12
Action
Limit
Pressure
2 / 12
Pressure Limit
90%
3 / 12
Pressure Mid
0%
4 / 12
Semi Auto Setup
On
Off
5 / 12
Set Limit
6 / 12
Sensor Position
49%
Open Settings
7 / 12
Action
Limit
Pressure
8 / 12
Pressure Limit
90%
9 / 12
Pressure Mid
0%
10 / 12
Semi Auto Setup
On
Off
11 / 12
Set Limit
12 / 12
Sensor Position
49%
1 / 12
Function Action (1/12) is shown highlighted. Use
to scroll through functions.
Functions will be highlighted in turn.
Summary of Contents for Skilmatic SI 3
Page 39: ...39...