- 10 –
______________________________________________________________________
in_hypalm_aw1_v3.doc
c)
AwQuick Mode:
Upon selecting this mode, the HygroPalm AW1 displays the following:
4
Q D w e l l [ m i n ]
Use the UP or DOWN arrow to set the dwell time. In the above example, the HygroPalm AW1 will
wait 4 minutes before processing the humidity data with the AwQuick algorithm.
Default value: 4 min (recommended value)
Press the
ENTER
key to record and confirm this setting. The display now shows the following:
0. 0 1
Q T S [
q
C / m i n ]
Use the UP or DOWN arrow to set the definition of stable temperature. In the above example, the
HygroPalm AW1 will consider temperature to be stable when the rate of change of the
temperature signal is less than 0.01
q
C per minute.
Default value: 0.01
q
C / min
Press the
ENTER
key to record and confirm this setting. The AwQuick mode is now active.
When ready to measure, press the ENTER key to access and start the AwQuick mode. Press the
ENTER key again to stop the AwQuick mode at any time and return to the Standard mode.
Notes:
When the HygroPalm AW1 is in the AwQuick mode, pressing the ENTER key automatically
starts the AwQuick mode, unless the MENU key was pressed immediately before pressing
the ENTER key.
The ON / OFF key is inoperative in the AwQuick mode. You must stop the AwQuick mode in
order to be able to turn the power of with the red ON/OFF key.
For instructions on using the AwQuick mode, see Water Activity Measurement.