© 2009-2010; Rotronic AG                                                                                                                             E-M-HP22-V1_21

 

5  

Maintenance 

5.1    Replacing the battery 

To replace the battery, turn the latching button counter-clockwise and pull out the battery holder.  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

5.2    Service cable 

  Cable AC3006 converts UART (service connector) to USB and is used to connect the HP22 to a USB port 

of a PC running the ROTRONIC HW4 software. Prior to using this cable, the ROTRONIC USB driver must 
be installed on the PC.  Both the driver and the installation instructions (document 

E-M-HW4v2-Main

) are 

located on the HW4 CD. 

  As an alternative, cable AC2001 is used to connect the HP22 to a probe input of the HP23 hand-held 

calibrator. For service purposes, the HP23 offers essentially the same functionality as the HW4 software. 

 

5.3    Location of the service connector (mini USB type) 

 
 

WARNING

: the service connector is a UART 

interface with a mini-USB connector type. Do not 
connect the service connector directly to the USB 
port of a PC or hub. 
 
The service connector (UART interface) can be 
accessed without opening the enclosure after 
removing the protective red round cover. 
 

 
 
 
 

Summary of Contents for HygroPalm HP22

Page 1: ...sersdorf Switzerland Document code Unit HygroPalm HP22 hand held indicator User Guide Instruction Manual Document Type Page 1 of 16 Document title 2009 2010 Rotronic AG E M HP22 V1_21 HygroPalm HP22 H...

Page 2: ...ings 5 3 3 Interaction between the HP22 and probe functions 7 4 Operation 8 4 1 Display 8 4 2 Keypad 8 4 3 Internal menu 9 4 4 Changing the parameters shown on the display 10 4 5 Changing the unit sys...

Page 3: ...nd 100 to 200 C 148 to 392 F The temperature operating range of the HP22 electronics is limited to 10 60 C Based on the ROTRONIC AirChip 3000 digital technology the HP22 indicator and HygroClip 2 prob...

Page 4: ...23 hand held calibrator This allows showing the data measured by the HP22 on the other instrument display and also allows using the other instrument internal menu to do for example a humidity and temp...

Page 5: ...d data can be downloaded to a PC with the HW4 software time stamped and viewed MEASUREMENT LOOP VALIDATION Functions Description Simulator mode Used to make the HP22 generate fixed values for the humi...

Page 6: ...djustment K Device write protection Disabled HP22 and probe Menu access from keypad Enabled Limit humidity output to 100 RH Enabled probe Out of limit value digital display alarm Disabled HP22 and pro...

Page 7: ...e name of the HC2 probe is not displayed by HW4 and is replaced with the HP22 Input Name Calculation X X Psychrometric calculation HP22 setting overrides HC2 probe setting Data refresh rate X When set...

Page 8: ...idity or dew frost point and the bottom line corresponds to temperature The display can be configured to show a trend indicator on each line increasing value decreasing value In the event of an alarm...

Page 9: ...irmware version Serial Nbr Serial number Address RS 485 address Name Device name User defined SensorTest Humidity sensor status Off Good SQ Tuned Bad Record On Off Data recording by the probe max 2000...

Page 10: ...dvice for measuring humidity The most common source of error when measuring relative humidity is a difference between the temperature of the probe and the temperature of the environment At a humidity...

Page 11: ...or to using this cable the ROTRONIC USB driver must be installed on the PC Both the driver and the installation instructions document E M HW4v2 Main are located on the HW4 CD As an alternative cable A...

Page 12: ...done prior to adjusting humidity o The calibration point is automatically deleted from the probe memory after an adjustment o Because the HP22 has no real time clock the date of the adjustment is not...

Page 13: ...roups Search for USB Masters After finding the HP22 with HW4 expand the device tree to see the HP22 functions Select Probe and Probe Adjustment For further instructions see HW4 manual E M HW4v2 A2 001...

Page 14: ...xing ratio by weight R Vapor concentration at saturation Dvs Vapor partial pressure E Vapor saturation pressure Ew Start up time and data refresh rate HP22 Start up time 3 s typical Data refresh rate...

Page 15: ...s the HW4 configuration software service cables calibration accessories and spare dust filters please see document E M HC2 accessories 9 Supporting documents Document File Name Contents E M HC2 Probes...

Page 16: ...6 Document title 2009 2010 Rotronic AG E M HP22 V1_21 E T HumiDefs Humidity Definitions Note All document file names have an extension corresponding to the document release number This extension is no...

Reviews: