4.3 Internal menu
Note
: Unauthorized access to the menu can be prevented by disabling the “display menu” setting
(use the HW4 software > Device Manager > Display)
Main Menu
Menu Items
Selections / Information
Notes
Device Settings
Units
Metric / English
Back Light
Key Press / On / Off
Display backlight mode
Contrast
LC display contrast adjustment
Trend
On / Off
Trend indication on the display
Device Information
Version
Firmware version
Serial Nbr
Serial number
Address
RS-485 address
Type
Device type
Name
Device name
User defined
Probe Information
Version
Firmware version
Serial Nbr
Serial number
Address
RS-485 address
Name
Device name
User defined
SensorTest
Humidity sensor status
Off / Good / SQ-Tuned / Bad
Record
On / Off
Data recording by the probe
(max. 2000 values)
Humidity Adjust
RefValue
Humidity reference value
± 0.1 %RH steps
Acquired
Number of cal. points in probe
memory
<Acquire>
Save cal. point to probe memory
<Delete>
Erases all calibration points
<Adjust>
Effect depends on number of
calibration points
Temperature Adjust
RefValue
Temperature reference value
± 0.1 ˚C steps
<Adjust>
1-point adjustment only (offset)
o
Record:
both the recording mode (start / stop and the log interval cannot be changed from the menu and
are as configured with the ROTRONIC HW4 software
o
SensorTest:
Off means that the humidity sensor has not been tested due to the configuration settings of
the test. For a description of the automatic humidity sensor test and drift compensation (SQ-tuning) see
documents
E-T-AC3000-DF-V1
and
E-M-HW4v2-F2-006