Step-by-step maintenance instructions
Shut off steam supply to sterilizer.
Caution: hot pipes.
Shut off cold water supply.
Make sure jacket pressure is 0 psi.
Disconnect service unions at:
• Cold water inlet
• Hot water inlet
• Tempered water outlet
Remove the WATER-MIZER from piping and
take to a service sink. Add water and, with pipe
caps or hands covering the hot water inlet and
tempered water outlet, shake the WATER-
MIZER to dislodge mineral deposits. Dump
mineral deposits and rinse.
WATER-MIZER
®
maintenance.
Maintenance
Maintenance
:
Maintenance instructions
6
Annual preventive maintenance on your WATER-MIZER
is recommended.
If severe mineral accumulation has occurred,
remove the split nut holding the sensor in place
and remove the thermal sensor from the WATER-
MIZER chamber. Gently clean the sensor with an
abrasive pad.
Do not twist the capillary tube
during removal or cleaning.
Remove the built-in
thermometer from the WATER-MIZER. Using an
abrasive pad clean any accumulated mineral.
Fill the WATER-MIZER with cold water. Reconnect
service unions, and open cold water supply valve
and steam supply valve to the sterilizer to resume
operation.
Do not run the WATER-MIZER dry –
excessive noise may occur.
The temperature of the effluent should be checked
as part of the current planned maintenance proce-
dures for the sterilizer which it is attached to. The
internal thermometer should be checked against a
calibrated external thermometer (or temperature
probe) annually.