E211: Unknown command from
Input-pool!
The communication was OK, but
an unknown command was
received from the Input-pool.
▪
Check RS232 cable between
Elevator board and Interrliner
Transport board.
▪
If there are I2C errors in the
ESR analyser unit: check the
I2C cable(s).
▪
If there are I2C errors in the
Interrliner: check the I2C
cable(s).
E212: Bad communication
between ESR-unit and Input-pool!
The ESR analyser unit did not
receive a good answer (first part)
after sending an inquiry to the
Input-pool.
▪
Check RS232 cable between
Elevator board and Transport
board.
▪
If there are I2C errors in the
ESR analyser unit: check the
I2C cable(s).
▪
If there are I2C errors in the
InteRRliner: check the
InteRRliner I2C cable(s).
E213: Bad communication
between ESR-unit and Input-pool!
The ESR analyser unit did not
receive a good answer (second
part) after sending an inquiry to
the Input-pool.
▪
Check RS232 cable between
Elevator board and Transport
board.
▪
If there are I2C errors in the
ESR analyser unit: check the
I2C cable(s).
▪
If there are I2C errors in the
Interrliner: check the I2C
cable(s).
E214: No response from Input-
pool!
The Input-pool did not response to
the ESR analyser unit after sending
a result to the Input-pool.
▪
Switching off the Interrliner
during run mode could cause
this error. This is normal.
▪
Check RS232 cable between
Elevator board and Transport
board.
▪
If there are I2C errors in the
ESR analyser unit: check the
I2C cable(s).
▪
If there are I2C errors in the
Interrliner: check the I2C
cable(s).
Appendix for Interrliner
MRN-164_2-EN Version 2.02 Interrliner
103