Check sensors
Vacuum pressure check
▪
Go to tab
Maintenance
->
Check sensor
. Select
Check Flow sensor
box.
Flow:
0980
± 60 Abs:
0320
±10
If the flow is not in range there might be a blockage in the vacuum flow line to the flow sensor.
Fill Stop sensor check
▪
Go to tab
Maintenance
->
Check sensor
. Select
Check Fill stop sensor
box.
Fill stop sensor FS 90..
140
..165
Diluter Start sensor check
▪
Go to tab
Maintenance
->
Check sensor
. Select
Diluter start sensor
box.
Diluter start sensor 400-
550
-700
Measure sensor check
▪
Go to tab
Maintenance
->
Check sensor.
Select
Check measure sensor
box.
Measure sensor MS
50
±10
Temperature sensor check
▪
Go to tab
Maintenance
->
Check sensor.
Select
Check Temperature sensor
box.
Temperature sensor TS [Room temperature]
Diluent flow sensor check
▪
Go to tab
Maintenance
->
Check sensor.
Select
Check Diluent flow sensor
box.
Press test. When test is finished, signal Down and signal Up must be green.
Separator check
▪
Go to tab
Maintenance
->
Check sensor.
Select
Check Separator sensor
box.
Separator sensor <200
600
>700
Inspect or replace sample probe or outer needle
Inspect needle condition. If necessary replace the sample probe or outer needle.
Check and clean instrument
Prepare disinfectant:
(if not already prepared).
This disinfectant is for cleaning of all external parts that are exposed to blood.
1. Go to tab
[Maintenance]
and perform the
[End-of-day wash]
procedure.
2. Check system for leakage.
▪ Inspect the peristaltic pump tubes and connections for leaks.
▪ Check that liquid does not run back after the pumps have stopped.
3. Clean the outer needle with disinfectant
4. Check tubing from the syringe for trapped air bubbles.
5. Check Diluent syringe for trapped air bubbles.
Work instruction Interrliner
MRN-164_2-EN Version 2.02 Interrliner
111