1. Select button OK.
2. The needle goes down and the process is started.
3. When all the pipettes are filled, the needle goes back to the home position.
4. After the Fill & clean process is finished, a notification is given to stop the procedure.
5. Remove the adapter from the needle assembly and press "Stop".
The instrument is ready for normal operation.
Sensor check
Vacuum pressure check
▪
Go to tab
Maintenance
->
Check sensor
. Select
Check Flow sensor
box.
Flow:
0980
± 60 Abs:
0320
±10
If the flow is not in range there might be a blockage in the vacuum flow line to the flow sensor.
Fill Stop sensor check
▪
Go to tab
Maintenance
->
Check sensor
. Select
Check Fill stop sensor
box.
Fill stop sensor FS 90..
140
..165
Diluter Start sensor check
▪
Go to tab
Maintenance
->
Check sensor
. Select
Diluter start sensor
box.
Diluter start sensor 400-
550
-700
Measure sensor check
▪
Go to tab
Maintenance
->
Check sensor.
Select
Check measure sensor
box.
Measure sensor MS
50
±10
Temperature sensor check
▪
Go to tab
Maintenance
->
Check sensor.
Select
Check Temperature sensor
box.
Temperature sensor TS [Room temperature]
Diluent flow sensor check
▪
Go to tab
Maintenance
->
Check sensor.
Select
Check Diluent flow sensor
box.
Press test. When test is finished, signal Down and signal Up must be green.
Separator check
▪
Go to tab
Maintenance
->
Check sensor.
Select
Check Separator sensor
box.
Separator sensor <200
600
>700
Work instruction Interrliner
122
MRN-164_2-EN Version 2.02 Interrliner