Exchange Waste Cassette and blotting washer
1. Disconnect the two tubes from the waste pump
cassette.
2. Press levers (at three o'clock and nine o'clock
positions) and pull at the same time.
3. Remove the old blotting washer ESRI090026
around the motor shaft.
4. Clean peristaltic pump motor shaft using a tissue
soaked in alcohol.
5. Place the new blotting washer
ESRI090026
.
6. Insert new waste pump cassette
ESRI090921
until it clicks into place.
7. Remove the protection caps on from the tubes.
8. Connect the two tubes from the waste pump
cassette on the luers at the steel plate under the
waste pump.
Exchange Pinch valve tube
Replace the pinch valve tube
ESRI010246
Inspect or replace sample probe or outer needle
A faulty or broken needle can cause a fill time-out error or a dilution error.
If necessary replace the sample probe or outer needle.
Check and clean instrument
Prepare disinfectant:
(if not already prepared).
This disinfectant is for cleaning of all external parts that are exposed to blood.
Work instruction Interrliner