Work instruction Number
251
Page
1 of
2
Purpose:
Fill and Clean
Safety:
Bio Hazard area
Instrument
: Interrliner
Revision:
001, 2019
Applicable for Software version 6.12 and up.
Starrsed Cleaning Agent preparation Interrliner ESR analyser unit: Fill
and clean:
This cycle takes about 90 minutes.
1. Fill the clean adapter EHST110907 with hot de-ionized water. (+/-
150 ml, 80 °C)
2. Add 15 ml Starrsed Cleaning Agent (QRR 010905) to the hot
water in the adapter.
3. Place the cap on the adapter and mix well.
4. Put the adapter with cleaning solution on the lower tube holder.
5. Select
Maintenance
tab,
Prime/Clean
, button
Fill and clean
.
Start Fill and clean procedure:
Work instruction Interrliner
128
MRN-164_2-EN Version 2.02 Interrliner