To start the Sample mode:
▪
Close the cover and press button “Sample mode”.
▪
The cover is automatically locked and sample mode is switched ON
▪
If button “Sample mode” is pressed while the cover is still open, an audible and visible warning
message is given, instructing the operator to close the cover. Movement of the mechanisms is halted until
the cover is closed and locked.
To get access to the tube handling mechanism, e.g. for maintenance or other necessary operator
intervention:
▪
Switch the “Sample mode” OFF by pressing the sample button.
▪
The sample sequence is aborted in a coordinated and safe way. Once movement of the tube handling
mechanisms has stopped, the cover is unlocked.
▪
In case of an error condition that requires operator interaction, the “Sample mode” is automatically
switched OFF and the cover is unlocked.
General safety information
22
MRN-164_2-EN Version 2.02 Interrliner