The above picture is an example of the
Sample
screen of the ESR analyser unit in the normal operation
mode. If the software is set to service mode is this indicated with a yellow bar next to the menu.
When the ESR analyser unit is running in the Service mode all kinds of settings can be changed and the
instrument will run with the changed settings. For instance, when ESR time is set to 12 minutes, the
Carousel will move according this time setting to be in time at the measure position.
Do not
perform normal sampling when Service mode is activated.
When the ESR analyser unit is running in the NORMAL MODE, the instrument uses the standard saved
settings. For instance the ESR time is set back to 60 minutes or 30 minutes according the used method.
The User Manual button is at the bottom of the menu. Click this button to open the Interrliner User manual.
5.2.1 - Main screen sections
1
Carousel
:
This is a graphical representation of the carousel. Pipettes are filled on position B.
When an ESR is required the carousel is moving to the Measure position (C). Each
pipette is washed at position D. On the display, the belt is also moving accordingly.
The numbers (A) next to the pipettes are the numbers on the pipette belt.
When a pipette is filled successfully, a blue dot marks the filled pipette. In case of a
failure, the pipette is marked with a flashing red dot.
All the sample information can be found in tab
History
Starrsed program
MRN-164_2-EN Version 2.02 Interrliner