5.3.5.2 - QC abnormal results (table)
The results from Starrsed Control level A are shown.
Display Results (table)
QC sample ID:
Read from the barcode. The original Starrsed Control barcode (=batch number)
Linked lab ID:
The Lab ID is given if it is linked to the Starrsed Control sample ID
Sampling date:
The date and time when the QC sample was aspirated.
Expiration date:
If the Starrsed Control expiry date is exceeded, it is not possible to continue with this QC sample. The
sample is not measured, but the failed attempt is logged in the table.
Expected ESR:
This is the temperature corrected mean value (incorporated in the Starrsed barcode) and the accepted range
of deviation. The applicable values for the acceptable range depend on the user setting.
ESR 60:
Uncorrected result from QC sample.
ESR 60
T.Corr
.:
Temperature corrected result from QC sample.
T(°C):
Temperature at which the sample was measured.
Starrsed program
38