QC result out of range!
▪
Perform a new QC sample, normal samples will be
finished
▪
If this error persists check/clean instrument
QC sample expired!
Use a new batch of Starrsed Control
It is not possible to link this Lab ID. Lab ID is already linked!
The "Linked QC IDs" table may only contain one link to a
particular Lab ID.
Consider changing
Automatically remove linked QC ID after
result
option to YES
Last QC result was out of range! Continuing could produce incorrect results! Do you still want to
continue?
Result of last QC sample was not within acceptable range. ▪
The last QC result should be evaluated by authorized
staff to decide whether the Interrliner may run patient
samples depending on the the nature of errors
▪
Press "Yes" to continue sampling without performing
a new QC, press "No" to return and take appropriate
action.
General errors
Barcode is not accepted
Barcode cannot be read
Data is incorrect
Check barcode
QC sample is not accepted and not performed
Starrsed Control ID is not known in LIMS.
Check barcode
QC result is not visible in QC History
A specific QC result cannot be found in the list of results. Check Lab-ID link
Deviating results
Systematic QC errors with a shift in control values (QC results are out of range)
Troubleshooting