13.6 - Reagents installation
Note: since november 2015 a 5 liter bottle (including a specific level sensor) is delivered for storage of
Deionized water.
Use only the Mechatronics bulk reagent cubitainers on the Interrliner. For storage of deionized water, a
plastic 5l bottle is delivered (including a specific level sensor and adapter).
1. Open the swing door and place the containers on the turntable.
2. Remove the cubitainer screw caps and pull the necks of the bottle packs out of the cardboard box.
3. Install the level sensors and spacers.
4. Add new reagent information in the software.
WARNING:
Wrongly placed pickup tubes causes incorrect results or instrument malfunction.
Make sure to place the appropriate level sensors in the cubitainers by checking the colored
labels/colored tubes matching the color codes on the cubitainers.
General Maintenance
92
MRN-164_2-EN Version 2.02 Interrliner