The sensors and the reagents have the following color codes:
Reagent
Color code
Starrsed Rinse solution
Green
Starrsed Saline
Yellow
Starrsed Diluent
Grey
Deionized water
Blue
Starrsed Disinfectant
White
Notes
Deionized water
Add one or two drops of Starrsed Saline to the deionized water to avoid
Reagent alarm
.
Starrsed Diluent
The solution should be discarded if it becomes turbid.
If the Diluent does become turbid, replace the Diluent container.
Use Level sensor ESRI110937 (Blue label) for the bottle with De-ionized water.
After each reagents change, the fluid system must be primed:
1. Select
Maintenance
->
Prime / Clean
.
2. Perform the applicable prime step to fill the relevant tubes with reagent and remove air.
13.7 - Cleaning Touch screen display
Any standard glass cleaner can be used to clean the touchscreen, but avoid products containing ammonia or
high concentrates (>70%) of alcohol or solvents.
Always spray the glass cleaner on a microfiber cloth or towel and then clean the touchscreen. Glass cleaner
sprayed directly on the monitor could possibly leak inside a non-sealed unit and cause damage.
Dirt and fingerprints do not affect the operation of a properly sealed touch display.