14.1 - Appendix - Error list Interrliner ESR analyser unit
last updated: 01-09-2014
Error
Extra explanation
Reason/Solution
E2: Communication error! (Board:
%s (%x), Command: %x, TWSR:
%x E: %d)
Communication lost after 3 retries
between Computer and Interrliner.
•
Power cable not connected
on the communication PCB
mounted on the back panel.
•
An I2C cable not connected
•
Serial cable not connected
•
No power on one of the PCB's
•
Short circuit or fault on one
of the PCB's
E3: Measure motor timeout!
Measure head motor did not move
or motor is blocked.
•
Measure head is not at the
Home position.
•
Check the Home sensor.
•
Motor is faulty.
•
Motor driver on drive board is
faulty.
E4: Sample probe not in top
position! (home)
Sample probe not back at Home
position after sampling a tube.
•
Check sample probe home
sensor.
•
Sample probe motor is faulty.
•
Sample probe motor driver on
needle board is faulty.
•
Sample probe is blocked.
E5: Duplicated ID !!
Sample rejected. Sample already in
carousel.
•
Wait until sample is measured
•
Check general settings (Check
for duplicate ID's)
E6: Program was not properly shut
down. Check settings before
continuing!
There is a possibility that changed
settings which were not saved to
disk are lost.
•
Program stopped and
computer needed to be reset.
•
Computer reset after power
failure.
E7: Outer needle motor position
error! Timeout! (piercing)
Outer needle did not go down
within a certain time limit.
•
Outer needle motor is faulty.
•
Outer needle motor driver on
needle board is faulty.
•
Outer needle is blocked.
E8: Fillnozzle not in fill position!
Fill nozzle did not reach the fill
position within a certain time limit.
•
Fill nozzle motor is faulty.
•
Fill nozzle motor driver on
nozzle board is faulty.
•
Fill nozzle is blocked.
Appendix for Interrliner