E37: NACK received from host
after sending 'Sample result
string'!
Did not receive ACK from Host
after sending 'Sample result string'
3 times.
See E30
E40: Position settings error.
Settings loaded from Eeprom.
Check settings before sampling!
Position settings in Eeprom do not
match settings saved to file.
Settings in Eeprom OK and loaded
from Eeprom. Check positions and
save settings.
•
Configuration file maybe
corrupted.
E41: Timeout settings error.
Settings loaded from Eeprom.
Check settings before sampling!
Timing settings in Eeprom does
not match settings saved to file.
Settings in Eeprom OK and loaded
from Eeprom. Check timeouts and
save settings.
•
Configuration file maybe
corrupted.
E104: Needle unit not in up
position!
Could not start the position motor,
because the outer needle or
sample probe is not its home
position (top).
•
Check outer needle home
sensor.
•
Check sample probe home
sensor.
•
Faulty Outer needle motor.
•
Faulty Sample probe motor.
•
Check if needles are blocked.
•
Faulty motor drivers on the
needle board.
E116-118
Quality Control Errors
E135: Elevator unit not
functioning!
Elevator unit did not respond to
main unit during start-up.
▪
No power on the Elevator
board.
▪
I2C cable(s) not connected.
▪
Bad I2C cable(s).
Note:
the boards in the ESR
analyser unit are connected in
series.
E136: Indexer unit not functioning! Indexer unit did not respond to
main unit during start-up.
▪
No power on the Indexer
board.
▪
I2C cable(s) not connected.
▪
Bad I2C cable(s).
Note:
the boards in the ESR
analyser unit are connected in
series.