5.5.1.2 - End-of-day-wash procedure
▪
End-of-day wash:
All pipettes will be washed once and needle, fill-nozzle and rinse-nozzle (wash station) are primed.
5.5.2 - Check sensors
When there is a sensor alarm, an alarm indicator is shown in the tab
Maintenance
.
All functions to check the status of the sensors are grouped under button
Check sensors
.
▪
Check Fill stop sensor: Click the Check button, the OK sign is shown if the sensor value is in range.
▪
Check temperature sensor: Value must be equal to the actual room temperature near the pipette belt.
The value can be set in tab
Settings
.
▪
Check Diluter start sensor: This sensor is only used in
. If the diluter does not start during
the aspiration, the status of this sensor must be checked.
Click the Check button, the OK sign is shown if the sensor value is in range.
▪
Check Diluent flow sensor: This sensor is only used in
. When activated, the indicator
Stand-by is activated and the indicator Flow is not. When the button Test is clicked, the indicator Flow
must be shown as "activated". After finishing the test, both indicators must be shown as "activated".
▪
Check Separator sensor: Click the Check button, the OK sign is shown if the sensor value is in range.
▪
Check Flow sensor: Click the Check button, the OK sign is shown if the sensor value is in range.
▪
Check Measure sensor:
Clean sensor first before executing this function.
Click the Check button, the OK sign is shown if the sensor value is in range.
Starrsed program
52
MRN-164_2-EN Version 2.02 Interrliner