1. Blood clots or rubber debris from the tube cap in the sample.
▪ Check the condition of the outer needle.
2. Filling procedure stopped by operator.
3. Insufficient sample volume.
▪ Should be at least 1.4 ml.
4. Faulty filling nozzle or filling nozzle O-ring.
▪ Check filling nozzle and O-ring.
5. Incorrectly adjusted sample probe depth.
▪ Check needle depth,
Settings
-
General settings
-
Sample probe depth
<default 5 mm>
6. No or poor vacuum.
▪ Check vacuum
Maintenance
-
Check sensors
-
Check flow sensor
11.8 - Hazy reports
"Hazy" reports are usually caused by build-up of proteins on the inner wall
of the pipettes. Another cause is growth of micro organisms in the diluter
system. It is extremely important that the system is kept sterile.
Erroneous haziness looks more like haemolytic plasma than 'normal' hazy
plasma. If there is an abnormal number of 'hazy' aspects, check the
dispenser system carefully and clean it if there is any doubt.
Symptoms for a contaminated instrument:
1. If more than
3… 5
out of
20
measurements are reported
HAZY.
2. Poor meniscus may be reported.
3. Incorrect results.
In case of hazy reports:
1. First run an extra
2. Check after a day's run if haziness is decreased. See
3. When there are still many reports, it is recommended to clean the
diluter system. See
WI-240 Cleaning the diluent system
Example of haziness
Troubleshooting
76
MRN-164_2-EN Version 2.02 Interrliner