1. The needle comes down.
2. Fill a container with 150 ml hot (80 °C) de-ionized water.
3. Add 15 ml cleaning agent. (QRR 010905)
4. Stir the prepared solution.
5. Put the container close to the needle location.
6. Push the silicon tube over the sample probe.
7. Press
Continue
.
8. The fill and clean process is started.
13.4 - Level 3 maintenance
Level 3 maintenance is level 4 maintenance adding the following extra's.
1. Replace the pinch valve tube
ESRI010246.
2. Replace the blue disk filter
QWLV040003.
3. Replace the Peristaltic waste pump cassette
ESRI090921
including the Blotting washer
ESRI090026.
General Maintenance
90
MRN-164_2-EN Version 2.02 Interrliner