3-2 Temperature Measurements 

 
1. Connect the plug of the thermocouple to the connector of the thermometer. 

2. Select the desired input mode and 

°

C/

°

F. 

3. Use the sensing point of thermocouple to measure the surface to be measured. 

4. Read the stable reading. 

5. "Warning:  Do not measure the surface if the potential exceeds 60Vd.c. or 24Vr.m.s." 
 

3-3 How to change to J-type mode 

 
1. Set the thermometer to "OFF" condition. 

2. Press the ON/OFF key and HOLD key at the same time. 

3. Unpress the ON/OFF key but keep the HOLD key depressed for at least 2 seconds and then release it, the  

    thermometer goes to the J-type mode. 

4. In this mode, the J symbol is displayed and the K symbol is blanked. 

5. To revert back to K-type mode, switch the instrument OFF and ON again. 

Summary of Contents for RS53 206-3744

Page 1: ...DIGITAL THERMOMETER OPERATOR S MANUAL 53 206 3744 55 206 3750 ...

Page 2: ...DIGITAL THERMOMETER OPERATOR S MANUAL RS53 206 3744 RS55 206 3750 ...

Page 3: ...2 ...

Page 4: ...tal Thermometer 2 K type Bead Thermocouple 3 Operator s Manual 4 Protective Holster 5 Belt Clip 1 2 Front Panel Refer to Figure 1 and the following numbered steps to familiarize yourself with the meter s front panel controls and connectors 1 Digital Display Window The digital display window has two 4 digit LCD readout maximum reading 9999 display ...

Page 5: ...ed to hold the numeric reading on main display When pressed the HOLD annunciator is displayed Conversions are made but the reading is not updated 4 Power ON OFF Switch Turns the thermometer on or off 5 Key This key is used to control the timer The timer is only enabled while the minor display is blanked The maximum reading is 99 99 When the first press is made the is displayed and the timer starts...

Page 6: ... readings To view the stored maximum reading press this key once the MAX symbol will appear To view the stored minimum reading press this key again the MIN symbol will appear Additional presses of this key toggle between MAX and MIN Press and hold this key for more than two seconds will disable this mode 8 T1 T2 Switch Selects the T1 T2 mode for measuring 55 only 9 T2 Switch Selects the T2 mode fo...

Page 7: ...6 2 1 3 4 5 6 7 8 9 10 Figure 1 ...

Page 8: ...y Two 4 digit Liquid Crystal Displays LCD with a maximum reading of 9999 Polarity Indication Automatic positive implied negative indicated Overrange Indication OL or OL Low Battery Indication is displayed when the battery voltage drops below operating voltage Reading Rate 1 sec reading nominal 53 only T1 or T2 mode 1 sec reading nominal 55 only T1 T2 mode 2 sec reading nominal 55 only ...

Page 9: ...ified accuracy C 18 C or 28 C Power Requirements 9V Alkaline battery Battery Life Alkaline 90 hours Dimensions H x W x D 160mm x 64mm x 26mm without holster 170mm x 74mm x 39mm with holster Weight including battery 300gms without holster 430 gms with holster Accessories 2 K type bead thermocouples 55 1 K type bead thermocouple 53 battery installed operator s manual and protective holster ...

Page 10: ...mometer 1 K type Measurement Range Resolution Accuracy Input Protection 200 C 100 C 0 1 C 0 3 reading 1 C 60V d c or 24 Vr m s 328 F 148 F 0 2 F 0 3 reading 2 F 99 9 C 999 9 C 0 1 C 0 1 reading 0 7 C 147 8 F 999 9 F 0 2 F 0 1 reading 1 4 F 1000 C 1370 C 1 C 0 3 reading 1 C 1000 F 2498 F 2 F 0 3 reading 2 F ...

Page 11: ...T2 reading 4 0 F Measurement Range Resolution Accuracy Input Protection 200 C 100 C 0 1 C 0 3 reading 1 1 C 328 F 148 F 0 2 F 0 3 reading 2 2 F 99 9 C 760 C 0 1 C 0 1 reading 0 8 C 147 8 F 999 9 F 0 2 F 0 1 reading 1 6 F 1000 F 1400 F 2 F 0 3 reading 2 F 60V d c or 24 Vr m s ...

Page 12: ... 399 2 F 2 Tolerances 2 2 C or 0 75 from 0 C to 204 C 2 2 C or 2 0 from 0 to 40 C 3 Wire Length 1m with miniature plug Teflon tape insulated B 50BJ bead thermocouple 1 Temperature Range 40 C to 204 C 40 F to 399 2 F 2 Tolerances 2 2 C or 0 75 3 Wire Length 1m with miniature plug Teflon tape insulated ...

Page 13: ...O AVOID ELECTRICAL SHOCK DO NOT USE THIS INSTRUMENT WHEN VOLTAGES AT THE MEASUREMENT SURFACE EXCEED 24V r m s AC OR 60V DC TO AVOID DAMAGE OR BURNS DO NOT MAKE TEMPERATURE MEASUREMENTS IN MICROWAVE OVENS 3 1 Preparation and Caution before Measurement 1 Before measurement warm up for at least 30 seconds after connecting the thermocouple to the thermometer 2 If the instrument is used near noise gene...

Page 14: ... measure the surface if the potential exceeds 60Vd c or 24Vr m s 3 3 How to change to J type mode 1 Set the thermometer to OFF condition 2 Press the ON OFF key and HOLD key at the same time 3 Unpress the ON OFF key but keep the HOLD key depressed for at least 2 seconds and then release it the thermometer goes to the J type mode 4 In this mode the J symbol is displayed and the K symbol is blanked 5...

Page 15: ...eactions that can damage the thermocouple BATTERY REPLACEMENT The meter is powered by a single 9V battery Refer to Figure 2 and use the following procedure to replace the battery 1 Turn the meter off Remove the thermocouple connector 2 Remove the holster 3 Position the meter face down Remove the screw from the battery cover 4 Remove the battery cover 5 Lift the battery from case top and carefully ...

Page 16: ...15 BATTERY REPLACEMENT Battery cover Screw 9V battery Battery connector Figure 2 ...

Page 17: ...16 HOW TO USE THE TILT STAND AND HOLSTER Swing the stand out for easier meter reading Hang on a nail at the workbench ...

Page 18: ...ospares Composants Rur Norman King BP 453 60031 Beauvais Cedex Tel 33 3 44 10 15 15 Fax 33 3 44 10 16 00 Italy RS Components S p A Via Cadorna 66 20090 Vimodrone Milano Tel 39 2 27 425 1 Fax 39 2 27 425 207 Germany RS Components GmbH Hessenring 13b 64545 Morfelden Walldorf Tel 49 6105 401 234 Fax 49 6105 401 100 ...

Reviews: