Calibrating Temperature:
1. Turn the power switch to 'ON'.
2. Set the temperature control buttons to 400°C (750°F), then wait until the
temperature stabilizes.
3. Remove the Calibration screw cover (CAL).
4. When the temperature stabilizes, use a straight-edge(-) screwdriver or small
plus(+) screwdriver to adjust the screw (marked CAL at the station) until the
tip thermometer indicates a temperature of 400°C(750°F). turn the screw
clockwise to increase the temperature and counter clockwise to reduce the
temperature.
5. Replace the Calibration screw cover (CAL)
Note: Temperature measurement is located in the middle point of about 3mm
inside the air nozzle.
.
Note:
Explanation of Symbols
A.
“SLP “
represents the temperature is lower than 100
℃
in standby mode,
and the handle is resting in the holder.
B.
“S-E”
represents that the switch turned on but the sensor of heat gun is
damaged.
Terms of use
1. The air outlet and its surrounding area may be very hot.
2. It is recommended not to put the heat gun on the work top surface always
return to its holder.
3. Please keep the air outlet unblocked.
4. The distance between the air outlet and the object shall be no less than 2mm.
5. Choose a suitable nozzle in accordance with working requirements. Different
nozzles may have different temperatures.
Fuse replacement
When a fuse is blown, replace with the same type of fuse. (see below picture)
1. Unplug the power cord from the power socket.
2. The fuse holder is located under the AC power socket, use the slotted (
–)
screwdriver to loosen the fuse holder.
EN-5