Operating Manual RS 1410 

 

 rspro.com

 

 

 

DANGER 

 

Incorrect area of application! 

In order to prevent erratic behaviour of the product, personal injury or property dam-
age, the product must be used exclusively as described in the chapter Description in 
the operating manual. 

–  Do not use in safety / Emergency Stop devices!

 

–  The product is not suitable for use in explosion-prone areas!

 

–  The product must not be used for diagnostic or other medical purposes on pa-

tients!

 

–  The product is not intended to come into direct contact with food. For measure-

ment in foods, samples must be taken and discarded after the measurement!

 

–  Not suitable for use with requirements on functional safety, e.g. SIL!

 

2.3 

Safety instructions 

 

 

NOTE 

 

This product does not belong in children's hands! 

2.4 

Intended use 

The product is designed for measuring the conductivity in liquids. The measuring cell is 
connected permanently. 

 

 

Summary of Contents for 1410-1002

Page 1: ...H87 0 01 DK2RS 2 0 Operating Manual EN DE RS 1410 1002 Stock number 205 0958 rspro com ...

Page 2: ...3 The product at a glance 7 3 1 RS 1400 series 7 3 2 Display elements 7 3 3 Operating elements 8 4 Measurement Basics 9 4 1 General information about conductivity measuring 9 4 2 Conductivity measurement 9 5 Operation and maintenance 10 5 1 Operating and maintenance notices 10 5 2 Battery 10 5 2 1 Battery indicator 10 5 2 2 Changing battery 10 6 Operation 12 6 1 Opening the configuration menu 12 6...

Page 3: ...Operating Manual RS 1410 page 3 of 20 9 Service 20 9 1 Manufacturer 20 ...

Page 4: ...ges or excerpts from this operating manual are prohibited The manufacturer assumes no liability for print errors 1 3 Further information _ Software version of the product V1 3 or later _ For the exact product name refer to the type plate on the rear side of the product _ NOTE For information about the software version press and hold the ON button to switch on the product for longer than 5 seconds ...

Page 5: ... the device or to the environment in case of non observance _ NOTE This symbol indicates processes which can have a direct influence on operation or can trigger an unforeseen reaction in case of non observance 2 2 Foreseeable misuse _ The fault free function and operational safety of the product can only be guaranteed if applicable safety precautions and the device specific safety instructions for...

Page 6: ...ble for use in explosion prone areas The product must not be used for diagnostic or other medical purposes on pa tients The product is not intended to come into direct contact with food For measure ment in foods samples must be taken and discarded after the measurement Not suitable for use with requirements on functional safety e g SIL 2 3 Safety instructions _ NOTE This product does not belong in...

Page 7: ...attery indicator Evaluation of the battery status _ Unit display Display of units or type of mode min max hold _ Main display Measurement of the current conductivity value or value for min max hold _ Auxiliary display Corresponding temperature value for the value shown in the main display If applicable alternating with the temperature compensation ...

Page 8: ...enu _ Up Down button Press briefly Display of the min max value Change value of the selected parameter Long press Reset the min max value of the current measure ment Both simultaneously Rotate display overhead display _ Function key Press briefly Freeze measurement Hold Return to measurement display Call up next parameter Long press 2s Open menu frozen measurement is displayed Close menu changes a...

Page 9: ...with conductivities varying widely make sure to properly rinse and shake dry measuring cell If conductivity measured is much higher or lower than expected this may be due to the electrode being soiled with non conducting or conducting foreign materials Measuring cell has to be cleaned with a watery soap solution When measuring media with low conductivities the electrode has to be stirred sufficien...

Page 10: ...n this manner If desired a gradient correction can be carried out for the product in order to further optimise the accuracy in a narrow range This is only necessary for normal use See Adjustment of the measuring input 5 2 Battery 5 2 1 Battery indicator _ If the empty frame in the battery display blinks the batteries are depleted and must be replaced However the device will still operate for a cer...

Page 11: ...g handling instructions before replacing batteries and follow them step by step If disregarded the product could be damaged or the protection from moisture could be diminished 1 Unscrews the Phillips screws A and remove the cover 2 Carefully replace the two Mignon AA batteries B En sure that the polarity is correct It must be possible to in sert the batteries in the correct position without using ...

Page 12: ...ter Values Meaning _ Input INP OND Measured variable conductivity SAL Measured variable salt content salinity TDS Measured variable total dissolved solids _ Factor for TDS CTDS 0 40 1 00 Conversion factor for TDS measurement _ Temperature compensation TCOR OFF Do not compensate conductivity measurement NLF Non linear function for natural water in accordance with EN 27888 ISO 7888 Ground water surf...

Page 13: ...c shut off 15 30 60 120 240 Automatic shut off after a selected time in minutes during which no buttons have been pressed _ Backlight L TE OFF Backlight deactivated 15 30 60 120 240 Automatic shut off of the backlight after a selected time in seconds during which no buttons have been pressed ON No automatic shut off of the backlight _ Display unit UN T Temperature display in C F Temperature displa...

Page 14: ...be adjusted with the zero point correction and the gradient correction If an adjustment is made you change the pre adjusted factory settings This is signalled with the T OF T SL or S L display text when the product is switched on 1 The product is switched off 2 Press and hold the Down button 3 Press the On Off button to switch on the product and open the Configuration menu Release the Down button ...

Page 15: ...e T SL 0 00 No gradient correction 5 00 5 00 Gradient correction in Gradient correction for the conductivity value S L 1 000 No gradient correction 0 800 1 200 Multiplier for the gradient correction Formula Zero point correction Displayed value measured value T OF Gradient correction C Display measured value T OF 1 T SL 100 Gradient correction F Display meas value 32 F T OF 1 T SL 100 32 F Gradien...

Page 16: ...r probe defect Measuring range exceeded or under cut Cable breakage Defective sensor or probe Measurement outside of the measuring range Send in for repair Send in for repair No display unclear characters or no re sponse when but tons are pressed Battery depleted System error Product is defective Battery depleted Error in the product Product is defective Replace battery Send in for repair BAT Batt...

Page 17: ...uring range is undercut Measurement too low Measuring cell defect Stay within allowable meas urement range Check the measuring cell Send in for repair SYS ERR System error Error in the product Switch product on off Replace batteries Send in for repair ...

Page 18: ... measured value 0 5 FS Temperature 0 3 C Measuring cycle approx 10 measurements per second Updating of the display approx 2 times per second Display 3 line segment LCD additional symbols illuminat ed adjustable white permanent illumination Additional functions Min Max Hold Adjustment Offset and gradient correction temperature Gradient correction conductivity Housing Break proof ABS housing Protect...

Page 19: ...ies BAT Auto power OFF function The device switches off automatically if this is activated Directives and standards The devices conform to the following Directives of the Council for the harmonisation of legal regula tions of the Member States 2014 30 EU EMC Directive 2011 65 EU RoHS Applied harmonised standards EN 61326 1 2013 Emission limits Class B Immunity according to Table 2 Additional error...

Page 20: ...Operating Manual RS 1410 rspro com 9 Service 9 1 Manufacturer If you have any questions please do not hesitate to contact us RS Components Limited Birchington Road Corby Northamptonshire NN17 9RS WEE GF0002ZR Contact ...

Page 21: ...H87 0 01 DK1RS 2 0 Betriebsanleitung EN DE RS 1410 1002 Produktnummer 205 0958 rspro com ...

Page 22: ... Bestimmungsgemäße Verwendung 6 3 Produkt auf einen Blick 7 3 1 Die RS 1400er Serie 7 3 2 Anzeigeelemente 7 3 3 Bedienelemente 8 4 Grundlagen zur Messung 9 4 1 Allgemeines zur Präzisions Leitfähigkeitsmessung 9 4 2 Leitfähigkeitsmessung 9 5 Betrieb und Wartung 10 5 1 Betriebs und Wartungshinweise 10 5 2 Batterie 10 5 2 1 Batterieanzeige 10 5 2 2 Batteriewechsel 10 6 Bedienung 12 6 1 Aufruf des Kon...

Page 23: ...Betriebsanleitung RS 1410 Seite 3 von 20 9 Service 20 9 1 Hersteller 20 ...

Page 24: ...hen oder Auszüge aus dieser Betriebsanleitung sind verboten Der Hersteller übernimmt keine Haftung bei Druckfehlern 1 3 Weiterführende Informationen _ Softwarestand des Produktes Ab V1 3 _ Die genaue Produktbezeichnung entnehmen Sie bitte dem Typenschild auf der Rück seite des Produktes _ HINWEIS Informationen zum Softwarestand erhalten Sie indem Sie beim Einschalten des Produkts die Eintaste für ...

Page 25: ...WEIS Symbol weist auf Vorgänge hin die bei Nichtbeachtung einen indirekten Einfluss auf den Betrieb haben oder eine nicht vorhergesehene Reaktion auslösen können 2 2 Vorhersehbare Fehlanwendungen _ Die einwandfreie Funktion und Betriebssicherheit des Produktes kann nur gewährleis tet werden wenn bei der Benutzung die allgemein üblichen Sicherheitsvorkehrungen sowie die gerätespezifischen Sicherhei...

Page 26: ...fährdeten Bereichen geeig net Das Produkt darf nicht für diagnostische oder sonstige medizinische Zwecke am Patienten verwendet werden Das Produkt ist nicht für direkten Kontakt mit Lebensmitteln ausgelegt Bei der Messung in Lebensmitteln sind Proben zu nehmen die nach der Messung ver worfen werden Nicht für die Anwendung mit Anforderungen an die funktionale Sicherheit geeig net z B SIL 2 3 Sicher...

Page 27: ...Batterieanzeige Bewertung des Batteriezustandes _ Einheitenanzeige Anzeige der Einheiten oder Art des Modus Min Max Hold _ Hauptanzeige Messwert des aktuellen Leitfähigkeit Wertes oder Wert für Min Max Hold _ Nebenanzeige Zugehöriger Temperaturwert zum angezeigten Wert in der Hauptanzeige Gegebenenfalls im Wechsel mit der verwendeten Temperaturkompen sation ...

Page 28: ...Auf Ab Taste Kurz drücken Anzeige des Min Max Wertes Wert des ausgewählten Parameters ändern Lang drücken Zurücksetzen des Min Max Wertes auf aktuellen Messwert Beide gleichzeitig Anzeige drehen Überkopfanzeige _ Funktions Taste Kurz drücken Messwert einfrieren Hold Rückkehr zur Messwertanzeige Nächsten Parameter aufrufen Lang drücken 2s Menü aufrufen eingefrorener Messwert wird ange zeigt Menü be...

Page 29: ...er zu spülen und gut auszuschleudern Wird eine unerwartet hohe bzw niedrige Leitfähigkeit gemessen so kann dies auf eine Verschmutzung der Elektroden mit nichtleitenden bzw leitenden Fremdstoffen zurück zuführen sein Die Messzelle muss dann ggf mit einer wässrigen Seifenlösung gesäu bert werden Bei Messungen in Lösungen mit geringer Leitfähigkeit ist für eine ausrei chende Anströmung der Elektrode...

Page 30: ...ktur durchgeführt werden um die Genauigkeit in einem engen Bereich weiter zu optimieren Dies ist nur für den normalen Gebrauch nicht notwendig Siehe Justieren des Messeinganges 5 2 Batterie 5 2 1 Batterieanzeige _ Blinkt in der Batterieanzeige der leere Rahmen so sind die Batterien verbraucht und müssen erneuert werden Die Gerätefunktion ist jedoch noch für eine gewisse Zeit gewährleistet Erschein...

Page 31: ...d befolgen Sie diese anschließend Schritt für Schritt Bei nicht Beachtung kann dies zu Beschädigungen des Produktes oder zur Beein trächtigung des Schutzes von Feuchtigkeit kommen 1 Die Kreuzschlitzschrauben A herausschrauben und den Deckel abziehen 2 Vorsichtig die beiden Mignon AA Batterien B wechseln Auf richtige Polarität achten Die Batterien müssen ohne Kraftaufwand in die korrekte Lage einge...

Page 32: ...sen Sie die Funktions Taste los _ Parameter Werte Bedeutung _ Input INP OND Messgröße Leitfähigkeit SAL Messgröße Salzgehalt Salinität TDS Messgröße Filtrattrockenrückstand _ Faktor für TDS CTDS 0 40 1 00 Umrechnungsfaktor zur TDS Messung _ Temperaturkompensation TCOR OFF Leitfähigkeitsmessung nicht kompensieren NLF Nichtlineare Funktion für natürliche Wässer nach EN 27888 ISO 7888 Grund Oberfläch...

Page 33: ... 240 Automatische Abschaltung nach ausgewählter Zeit in Minuten wenn kein Tastendruck erfolgt _ Hintergrundbeleuchtung L TE OFF Hintergrundbeleuchtung deaktiviert 15 30 60 120 240 Automatische Abschaltung der Hintergrundbeleuch tung nach ausgewählter Zeit in Sekunden wenn kein Tastendruck erfolgt ON Keine automatische Abschaltung der Hintergrund beleuchtung _ Anzeigeeinheit UN T Temperaturanzeige ...

Page 34: ...gskorrektur kann der Temperatureingang justiert werden Wird eine Justierung vorgenommen so ändern Sie die voreingestellten Werkseinstellungen Dies wird beim Einschalten des Produktes mit dem Anzeigetext T OF T SL oder S L signalisiert 1 Das Produkt ist ausgeschaltet 2 Halten Sie die Ab Taste gedrückt 3 Drücken Sie die Ein Aus Taste um das Produkt einzuschalten und das Menü Konfiguration aufzurufen...

Page 35: ...L 0 00 Keine Steigungskorrektur 5 00 5 00 Steigungskorrektur in Steigungskorrektur für Leitfähigkeitswert S L 1 000 Keine Steigungskorrektur 0 800 1 200 Multiplikator für die Steigungskorrektur Formeln Nullpunktkorrektur Angezeigter Wert gemessener Wert T OF Steigungskorrektur C Anzeige gemessener Wert T OF 1 T SL 100 Steigungskorrektur F Anzeige gem Wert 32 F T OF 1 T SL 100 32 F Steigungskorrekt...

Page 36: ...ter Sensor oder Fühler Messung außerhalb zulässigem Bereich Zur Reparatur einschicken Zur Reparatur einschicken Keine Anzeige wirre Zei chen oder keine Reak tion auf Tastendruck Batterie verbraucht Systemfehler Produkt defekt Batterie verbraucht Fehler im Produkt Produkt defekt Batterie ersetzen Zur Reparatur einschicken BAT Batterie verbraucht Batterie verbraucht Batterie ersetzen ERR 1 Messberei...

Page 37: ...Betriebsanleitung RS 1410 Seite 17 von 20 SYS ERR Systemfehler Fehler im Produkt Produkt ein aus schalten Batterien tauschen Zur Reparatur einschicken ...

Page 38: ...etzt werden Genauigkeit Leitfähigkeit 0 5 v MW 0 5 FS Temperatur 0 3 C Messzyklus ca 10 Messungen pro Sekunde Aktualisierung der Anzeige ca 2 mal pro Sekunde Display 3 zeiliges Segment LCD zusätzliche Symbole beleuchtet weiß Leuchtdauer einstellbar Zusätzliche Funktionen Min Max Hold Abgleich Offset und Steigungskorrektur Temperatur Steigungskorrektur Leitfähigkeit Gehäuse bruchfestes ABS Gehäuse ...

Page 39: ...bei verbrauchter Batterie BAT Auto Power Off Funktion falls aktiviert schaltet sich das Gerät automatisch ab Richtlinien und Normen Die Geräte entsprechen folgenden Richtlinien des Rates zur Angleichung der Rechtsvorschriften der Mitgliedstaaten 2014 30 EU EMV Richtlinie 2011 65 EU RoHS Angewandte harmonisierte Normen EN 61326 1 2013 Störaussendung Klasse B Störfestigkeit nach Tabelle 2 Zusätzlich...

Page 40: ...Betriebsanleitung RS 1410 rspro com 9 Service 9 1 Hersteller Wenn Sie einmal Fragen haben zögern Sie nicht uns zu kontaktieren RS Components Limited Birchington Road Corby Northamptonshire NN17 9RS WEE GF0002ZR Kontakt ...

Reviews: