7. Specifications
Heavy Duty True RMS Digital Multimeter / English
16/12/2017 Version No. 001
Function
Range
Resolution
Accuracy
DC Voltage
600 mV
0.1 mV
±(0.8% r 5 digits)
6 V
0.001 V
±(0.5% r 3 digits)
60 V
0.01 V
600 V
0.1 V
1000 V
1 V
±(0.8% r 5 digits)
AC Voltage
45 Hz to 1 kHz
600 mV
0.1 mV
±(1.5% r 8 digits)
6 V
0.001 V
60 V
0.01 V
600 V
0.1 V
1000 V
1V
±(1.2% r 5 digits)
All AC voltage ranges are specified from 5% of range to 100% of range
Function
Range
Resolution
Accuracy
DC Current
600 μA
0.1 μA
±(1.0% r 3 digits)
6000 μA
1 μA
60 mA
0.01 mA
600 mA
0.1 mA
6 A
0.001A
±(1.5% r 3 digits)
10 A
0.01 A
(20 A: 30 seconds max with reduced accuracy)
AC Current
45 Hz to 1 kHz
600 μA
0.1 μA
±(1.5% r 3 digits)
6000 μA
1 μA
60 mA
0.01 mA
600 mA
0.1 mA
6 A
0.001 A
±(2.0% r 5 digits)
10 A
0.01 A
(20 A: 30 seconds max with reduced accuracy)
All AC Cur rent ranges are specified from 5% of range to 100% of
range
NOTE:
Accuracy is stated at 18°C to 28°C (65°F to 83°F) and less than 75% RH
Function
Range
Resolution
Accuracy
Resistance
600 Ω
0.1 Ω
±(1.2% r 5 digits)
6k Ω
0.001 kΩ
60 kΩ
0.01 kΩ
600 kΩ
0.1 kΩ
6M Ω
0.001 MΩ
60 MΩ
0.01 MΩ
±(2% r 20 digits)
Heavy Duty True RMS Digital Multimeter / English
16/12/2017 Version No. 001
6-2. Replacing the Fuses
WARNING:
To avoid electric shock, disconnect the test leads from any source of
voltage before removing the fuse cover.
WARNING:
To avoid electric shock, do not operate your meter until the fuse cover is
in place and fastened securely.
1. Disconnect the test leads from the meter.
2. Remove the battery cover.
3. Gently remove the old fuse and install the new fuse into the holder.
4. Always use a fuse of the proper size and value (0.8 A/1000 V fast blow for the 600 mA
range [SIBA 70-172-40], 10 A/1000 V fast blow for the 20 A range [SIBA 50-199-06]).
5. Replace and secure the rear cover, battery and battery cover.