Resistance and Continuity test
Frequency (electronic circuits)
Diode test
Range
Resolution
Accuracy
Buzzer
Protection against
overcharge
600.0Ω
0.1Ω
±(1.2%r 10dgt)
<50Ω
600VDC/ACrms
6.000kΩ 0.001kΩ
±(1.2%r 5digits)
60.00kΩ 0.01kΩ
600.0kΩ 0.1kΩ
6.000MΩ 0.001MΩ
60.00MΩ 0.01MΩ
±(2.5%r 10dgt)
Range
Resolution
Accuracy
Protection against
overcharge
40.00Hz to 100kHz 0.01Hz to 0.001kHz
±(0.5%reading) 600VDC/ACrms
Function
Test current
Max voltage with open circuit
<1.5mA
3.0VDC
AC Current ( with Current Transducer )
Range
Transducer ratio Resolution
Accuracy (*)
Protection against
overcharge
(50Hz to 60Hz)
30A*
100mV/1A
0.01A
±(1.2%rdg +
10digits)
1000VDC/ACrms
300A*
10mV/1A
0.1A
3000A*
1mV/1A
10A
(*) Accuracy specified from 5% to 100% of the measuring range
(*)Do not include the accuracy of current Transducer
* with current sensor 320B( 30A*/300A*/3000* Rang Corresponding DT-320B Rang)
Distorted ,pulsed , triangle or trapezia waveform Accuracy: ±(10%rdg + 10dgt)
Sensitivity: 2Vrms
Professional True RMS Digital Multimeter/ English
Duty Cycle
Range
Resolution
Accuracy
20% to 80%
0.1%
±(1.2%r 2digits)
Pulse frequency range: 40Hz to 100kHz, Pulse amplitude: ±5V (100µs to 100ms)
Capacity
Temperature with K-type probe
Range
Resolution Accuracy
Protection against overcharge
999.9nF
0.1nF
±(1.2%r 8digits)
600VDC/ACrms
9.999µF
0.001µF
±(1.5%r 8digits)
999.9µF
0.1µF
±(1.5%r 8digits)
99.99mF
0.01 mF
±(2.5%r 20digits)
Range
Resolution Accuracy (*)
Protection against
overcharge
-40.0°C to 600.0°C 0.1°C
±(1.5%r 3°C)
600VDC/ACrms
600°C to 1000°C
1°C
-40.0°F to 600.0°F 0.1°F
±(1.5%rdg+ 5.4°F)
600°F to 1832°F
1°F
(*) Instrument accuracy without probe ; Specified accuracy with stable environmental
temperature at ±1°C
For long-lasting measurements, reading increases by 2°C
Professional True RMS Digital Multimeter/ English