background image

Instruction Manual 
ICM A6N  
Clamp Meter     

EN

 

DE

 

ES

 

IT

 

FR

 

 

1000Vdc

CAT

750Vac

CAT

600V

Summary of Contents for ICM A6N

Page 1: ...Instruction Manual ICM A6N Clamp Meter EN DE ES IT FR 1000Vdc CAT 750Vac CAT 600V ...

Page 2: ...r to Instruction Manual Alternating Current ac Battery Earth Equipment protected throughout by double or reinforced insulation Conforms to EU directives Dispose of this equipment in accordance with local regulations E Application around and removal from hazardous Live conductors is permitted WARNING Examine the instrument and probes and leads before use Do not use the instru ment if it is wet or d...

Page 3: ... Disconnect the circuit power and discharge all high voltage capacitors before making resistance continuity or diode measurements Do not use the instrument in a Hazardous Area or around explosive gasses or vapours Wear suitable Personal Protective Equipment when working around or near hazardous Live conductors which could be accessible Caution Disconnect the test leads from the test points before ...

Page 4: ...fied personnel This instrument should be calibrated yearly or more frequently if used in harsh conditions or if it is suspected of being inaccurate For calibration and repair contact RS Components the address is given at the end of these instructions Cleaning Periodically wipe the case with a damp cloth and detergent Do not use abrasives or solvents ...

Page 5: ...capacity of the battery Replace the battery when 0 is shown or the low battery symbol remains visible in the display After a successful battery self test the instrument will automatically select resis tance Ω continuity mode Auto Hold Enable Disable Auto Power Off Enable Disable ...

Page 6: ... for the jaw Barrier for Hand Guard Do not hold the meter above the Barrier Do not attempt to measure while power ing up as it will cause the self test to fail Caution If FAIL is shown in the display do not use the instrument as an internal fault has occurred AC Current measurement ...

Page 7: ...vice connect the black lead before the red lead When removing the test leads remove the red lead before the black lead Input Impedance Note The input impedance is 4K for an input voltage up to 30V and increases automatically with input voltage to approxi mately 375KΩ at 750V ...

Page 8: ...EN 8 Continuity Resistance Diode ...

Page 9: ...satisfies the following conditions Resistance Between 0 and Ω ACV Between 0 and 0 9V DCV Between 0 4 and 0 2V DCV Between 1 0 and 2 0V The LCD display shows diode mode when the voltage at the input terminals satisfies the following conditions DCV Between 0 4 and 0 8V The LCD display shows AC current A mode when the current sensed by the jaws satisfies the following conditions ACA Between 0 9A and ...

Page 10: ...ons are met in Data hold mode 1 The instrument measures a characteristic different from the LCD reading e g a voltage reading is held on the display and an attempt is made to measure an AC current 2 The measured signal is the same characteristic as the LCD reading but is 50 counts or more greater than the held LCD reading Auto Hold ACA only ...

Page 11: ...ttery indicator appears on the LCD replace the batteries with the type given in the specifications WARNING Disconnect the test leads from the circuit and the instrument before removing the battery cover Refer to the following figure to replace the batteries ...

Page 12: ...y Calibration One year calibration cycle Operating temperature 0 C to 30 C 32 F to 86 F 80 RH 30 C to 40 C 86 F to104 F 75 RH 40 C to 50 C 104 F to 122 F 45 RH Storage temperature 20 to 60 C 4 F to 140 F 0 to 80 RH with batteries removed from the instrument Measurement Category Installation Category per IEC 61010 1 2001 CAT Ⅲ 600V Measurement Category l is for measure ments performed on circuits n...

Page 13: ...ement Category lV is for measure ments performed at the source of the low voltage installation Examples include measurements on primary overcurrent protection devices and electricity meters Conductor Size 27mm diameter maximum Pollution degree 2 Operating altitude 2000m 6562 ft EMC EN 61326 1 Shock vibration Sinusoidal vibration per MIL T 28800E 5 to 55 Hz 3g maximum Drop Protection 4 feet drop on...

Page 14: ...eases with input voltage to approximate 375KΩ at 750V Resistance Continuity Diode Function Range Accuracy Ω 0 0Ω 9999Ω 2 2 dgt 0 4V 0 8V 1 3 dgt Overload protection 1000 VDC 750 VAC Max open circuit voltage 2 0V Continuity check Internal sounder activates if the resistance of the circuit under test is less than 25Ω It will then turn off if the resistance is increased beyond 400Ω Specified from 0 C...

Page 15: ...tion contamination or abnormal conditions of operation or handling Any implied warranties arising out of the sale of this product including but not lim ited to implied warranties of merchantability and fitness for a particular purpose are limited to the above RS Components shall not be liable for loss of use of the instru ment or other incidental or consequential damages expenses or economic loss ...

Page 16: ...EN 16 ...

Page 17: ...n Road Shanghai 200001 China www rs components com Europe RS Components Ltd PO Box 99 Corby Northants NN17 9RS United Kingdom www rs components com Japan RS Components Ltd West Tower 12th Floor Yokohama Business Park 134 Godocho Hodogaya Yokohama Kanagawa 240 0005 Japan www rs components com U S A Allied Electronics 7151 Jack Newell Blvd S Fort Worth Texas 76118 U S A www alliedelec com South Amer...

Reviews: