IDM-8351 Instruction Manual/English
164
17/11/2016 Version No. 002
100.000mV
0.001mV
119.999
1% + 100
0.3% + 100 1.5% +300
5% + 300
1.00000 V
0.00001V
1.19999
1% + 100
0.2% + 100
1% +100
3% + 200
10.0000 V
0.0001V
11.9999
1% + 100
0.2% + 100
1% +100
3% + 200
100.000 V
0.001V
119.999
1% + 100
0.2% + 100
1% +100
3% + 200
750.00 V
[3]
0.01V
765.00
1% + 100
0.2% + 100
1% +100
3% + 200
[1] Specifications are for sine wave inputs that are greater than 5% range.
[2] Rate in Fast, Input ACV Frequency > 200Hz.
[3] The specifications are guaranteed to an input of 750V. A beeping alarm will go off when
the input value is higher than 750V.
AC Current
Range
[1][3]
Resolution
Full Scale
Accuracy
20 Hz to 45 Hz 45 Hz to 2 kHz 2 kHz to 10kHz
10.0000mA 0.0001mA
11.9999
1.5% + 100
0.5% + 100
2% + 200
100.000mA
0.001mA
119.999
1.5% + 100
0.5% + 100
2% + 200
1.00000A
0.00001A
1.19999
1.5% + 100
0.5% + 100
2% + 200
[2]
10.0000A
[4]
0.0001A
11.9999
1.5% + 100
1% + 100
-
[1] Specifications are for sine wave inputs that are greater than 5% of range.
[2] Input current (5k ~ 10kHz)<220mArms.
[3]
The accuracy of ACI+DCI is equal to ACI’s with 10 more digits added.
[4] The specifications are guaranteed to 10A. A beeping alarm will go off when the input
current being measured is higher than 10A.
Frequency Accuracy
Rate
10Hz to 1MHz
[1]
Slow (>10Hz)
0.01% + 3
Med (>20Hz)
Fast (>200Hz)