Set Continuity Threshold
The continuity threshold defines the maximum resistance allowed in the DUT when
testing the continuity.
Range
Threshold
0 to 1000
Ω (Default Threshold:10Ω)
Resolution
1Ω
Procedure
1. Press MENU.
2. Go to the MEAS menu on level 1
3. Go to the CONT menu on level 2
4. Set the continuity threshold level in ohms.
5. Press the Enter key to confirm the continuity settings.
6. Press EXIT to exit the CONT setting menu.