IDM-8351 Instruction Manual/English
56
17/11/2016 Version No. 002
Set the Reference Junction Temperature
When a thermocouple is connected to the DMM, the temperature difference
between the thermocouple lead and the DMM input terminal should be taken into
account and be cancelled out; otherwise an erroneous temperature might be
added. The value of the reference junction temperature should be determined by
the user.
Range
SIM
0 ~ 50˚C (default: 23.00˚C)
Resolution
0.01˚C
Procedure
1. Press the MENU key.
2. Go to TEMP on level 1.
3. Go to SIM on level 2.
4. Set the SIM (simulated) reference junction temperature.
5. Press the Enter key to confirm.
6. Press the EXIT key to exit from the temperature menu.
Display
Reference junction
temperature setting
SIM menu
indicator