22
Electrical Specifications
(1) Test Frequency :
Range
Resolution
Accuracy
100.00 Hz
0.01 Hz
± 0.01%
120.00 Hz
0.01 Hz
± 0.01%
1.0000 kHz
0.1 Hz
± 0.01%
10.000 kHz
1 Hz
± 0.01%
100.00 kHz
10 Hz
± 0.01%
(2) Test Signal :
AC Signal Level :
600mVrms
AC Signal Accuracy :
±10%
DC Bias Level :
1V
DC Bias Accuracy :
±10%
(3) Test Cable :
Model
Length
Bandwidth
Type
SMD Test Probe
60cm
1MHz
5-Wire
4-Wire Test Probe
60cm
1MHz
5-Wire
Alligator Clip Set
15cm
1kHz
2-Wire
Accuracy :
± (A x B) (% of reading)
A : Basic Accuracy as specified by
B : Test Cable Accuracy
B(%) = 1 + (L x F x T)
L(m) : Cable Length
F(MHz) : Test Frequency
T : Cable Type. If the cable is 5-Wire type, the "T" is 40, and the
other is 4,000.
When measuring by basic accuracy that following
conditions must be met :
1. Ambient temperature : 23°C ± 5°C < 80%RH.
2. Test cable length : 0 m
3. Open and short corrections have been performed.
4. D ≤ 0.1 for C, L measurements; Q ≤ 0.1 for R measurements.
See the operation manual for additional conditions.