26
(7) DCR
Range
Resolution
Accuracy
200.00
Ω
0.01
Ω
± (0.2% + 5d)
2.0000K
Ω
0.0001K
Ω
20.000K
Ω
0.001K
Ω
200.00K
Ω
0.01K
Ω
2.0000M
Ω
0.0001M
Ω
20.000M
Ω
0.001M
Ω
± (0.5% + 5d)
200.00M
Ω
0.01M
Ω
± (1.0% + 5d)
[1]
[1] < 50dgt rolling.
Input Protection
:
30V
DC
or 30V
AC
rms
Minimum Resolution
:
0.01
Ω
in the 200.00
Ω
range.
Measuring Time
:
2 seconds
2-wire Accuracy
:
Add 1.0% to accuracy.
(8) D & Q
Definition Q=1/D
Range
:
2.000 ~ 2000
Minimum Resolution
:
0.001 in the 2.000 range.
Accuracy
:
Accuracy of Main Reading x (1+D)
Input Protection
:
30V
DC
or 30V
AC
rms
2-wire Accuracy
:
Add 1.0% to accuracy.
(9) ESR
:
The specification of ESR is same as Resistance.
(10)
θ
:
Range
Resolution
Accuracy
-90.0° ~ 90.0°
0.1°
± (0.2% + 5d)
Input Protection
:
30V
DC
or 30V
AC
rms
2-wire Accuracy
:
Add 1.0% to accuracy.