3.9 Auto Calibration
Whenever users move the sliding switch (RTD 4W, 3W, 2W, TC), enter and
exit from SETUP, or press M/S button to change MEASURE or SOURCE
mode, the calibrator will always re-calibrate itself to ensure the accuracy.
Users shall see the text of “CAL..1 …” displayed at the left bottom corner of
LCD.
At this moment, the calibrator does not response to any input until calibration
is done.